EN
  • ISO 11443
  • ASTM D3835

Capillary Rheometer | Capilograph™ | 583

รุ่นNo. 583

Toyo Seiki · Melt Flow Index (MFR, MVR) / Melt Viscosity

The Capilograph measures melt viscosity of polymers by detecting the shear rate and shear stress of melt polymer when it flows out of the capillary die. Die swell and melt strength can also be measured by equipping it with optional attachments. The twin bore barrel model (F-2) allows easy Bagley & Weissenberg-Rabinowitsch correction (for determination of true viscosity) or elongational viscosity calculation.

เอกสารดาวน์โหลด

ข้อมูลจำเพาะ

ISO 11443Plastics — Determination of the fluidity of plastics using capillary and slit-die rheometers
ASTM D3835Standard Test Method for Determination of Properties of Polymeric Materials by Means of a Capillary Rheometer

ตารางเปรียบเทียบรุ่น

ItemValue
ISO 11443Plastics — Determination of the fluidity of plastics using capillary and slit-die rheometers
ASTM D3835Standard Test Method for Determination of Properties of Polymeric Materials by Means of a Capillary Rheometer

เลื่อนตารางไปทางข้างเพื่อดูทุกคอลัมน์

– Safety door is included as standard.
ModelF-1F-2
Temperature range60 to 400°C (500°C optional)
Temperature accuracy±1°C (Less than 300°C) / Within ±2°C (300°C or more)
Number of boresSingleTwin
Bore diameterSelection: / • ∅9.55mm / • ∅15mm∅15mm
Barrel lengthLength: 350mm, effective length: 250mm
Piston speed0.1 to 1500mm/min.
Force range• 25kN max (bore dia. ∅9.55mm) / • 50kN max (bore dia. ∅15mm)25kN max (bore dia. ∅15mm)
Power supplyThree-phase, AC200V, 50/60Hz, 5.2kVA (Main unit)
Compressed air requirement0.5MPa
DimensionsW1060 × D1100 × H1950 to 2360mm
WeightApprox. 650kg

เลื่อนตารางไปทางข้างเพื่อดูทุกคอลัมน์

วิดีโอ