- IEC 60068
- IEC 60749
- MIL-STD-202F
Altitude Chamber
LIB · Altitude, Vibration & Combined Simulation
LIB altitude chamber can simulate the air pressure at an altitude below 40,000 meters, and is used for combined temperature altitude humidity testing. The altitude test chamber can carry out low-pressure tests independently, and can also carry out comprehensive tests high temperature, low temperature and humidity. The high altitude test chamber is used to test the stability of the products when it is subjected to altitude, temperature, humidity, and comprehensive effects in the air or at high altitudes.
Datasheets
- Lib Th Chamber Cataloglib-th-chamber-catalog.pdfDownload PDF
- Lib Industry Temperature Humidity Chamber User Manuallib-industry-temperature-humidity-chamber-user-manual.pdfDownload PDF
- Lib Industry Temperature Humidity Chamber Applicable Standardslib-industry-temperature-humidity-chamber-applicable-standards.pdfDownload PDF
- Lib Industry Temperature Humidity Chamber Maintenance Processing Listlib-industry-temperature-humidity-chamber-maintenance-processing-list.pdfDownload PDF
- User Manual For Temperature Humidity Chamberuser-manual-for-temperature-humidity-chamber.pdfDownload PDF
Specification
| Internal dimensions | D 1500 W 1000 H 970 mm |
|---|---|
| External dimensions | D 2050 W 1300 H 2200 mm |
| Temperature range | —40℃ to +150℃ |
| Humidity range | 20% ~ 95% (more than 50Kpa can be controlled) |
| Temperature fluctuation | ≤ ±0.5℃ |
| Temperature deviation | ≤ ±2.0℃ |
| Temperature uniformity | ≤ 2.0℃ (under normal pressure and no load) |
| Heating rate | 3℃/min |
| Cooling speed | 2℃/min |
| Pressure range | Atmospheric pressure to 0.5kPa |
| Pressure deviation | ± 2Kpa |
| Depressurization time | Atmospheric pressure to 0.5k ≤30min |
| Pressure enclosure | Manganese steel, surface spray treatment |
| Other outer wall materials | Double-sided galvanized steel plate, surface spray treatment |
| Inner wall material | SUS304 stainless steel plate |
| Vacuum system | Vacuum pump, air pumping regulating valve, inflation regulating valve |